Touch screen intelligent digital multimeter
VΩ
directory
directory
1. Safety information............................................................... 1
1.1 safety instructions................................................................ 1
1.2 electrical symbols................................................................. 1
1.3 maintenance..........................................................................2
2. The description......................................................................2
2.1 part name.............................................................................. 2
2.1 LCD touch monitor................................................................3
3. The technical indexes........................................................... 4
3.1 comprehensive index........................................................... 4
3.2 technical indicators.............................................................. 5
4. Operating instructions........................................................ 8
4.1 keep reading......................................................................... 8
4.2 the flashlight......................................................................... 8
4.3 automatic shutdown............................................................ 8
4.4 measurement........................................................................ 8
4.5 non-contact voltage detection (NCV)................................9
4.6 dc voltage measurement..................................................... 9
4.7 ac voltage measurement..................................................... 10
4.8 ac/dc current measurement................................................ 11
4.9 resistance measurement......................................................11
4.10 wire test............................................................................... 12
4.11 capacitance measurement................................................ 12
4.12 temperature measurement............................................... 12
4.13 diode test.............................................................................13
4.14 circuit on-off test................................................................13
5. Maintain..................................................................................14
5.1 Type - c line charging........................................................... 14
5.2 replacement pens and..........................................................14
6. The attachment......................................................................14
Safety information
1.Safety information
To fully understand the function of the meter and ensure
safe operation, please read and follow it carefully Follow the
use method of this manual.
Special care should be taken when using this meter,
improper use may cause electric shock or damage to the
meter. In use, it should follow the usual safety regulations
and take effective safety precautions.
The instrument meets the safety requirements of IEC-61010-1,
IEC-61010-2-030, IEC-61010-2-032) electronic measuring
instrument, belongs to secondary pollution, and the overvoltage
standard is CAT III 600V
Please follow the safety operation guide and use the meter
safely.
1.1Safety instructions
1.1.1 When using the meter, users must follow standard safety rules
' -Universal protection against electric shock
' -Prevent misuse of meter
1.1.2 After receiving the meter, check for damage in transit.
1.1.3 After storage and shipment in poor condition, check and
confirm whether the meter is damaged.
1.1.4 The pen must be in good condition. Before use, check whether
the insulation of the pen is damaged and whether the metal
wire of the wire is exposed
1.2Electrical symbols
Note (For important safety information, please refer to the
instruction manual)
Can be used on dangerous charged objects
Double insulation protection (Class II)
In accordance with iec-61010-1 overvoltage (installation)
Class II, contamination level 2 refers to the level of voltage
protection provided to withstand pulses.
describe
Comply with European Community (EU) standards.
Grounding
1.3maintenance
1.3.1Do not attempt to open the bottom case to adjust or repair
the instrument. Such operations should only be performed
by technicians who are fully aware of the instrument and the
risks of electric shock.
1.3.2The pen should be removed from the line under test before
opening the instrument backing.
1.3.3To avoid electric shocks that may be caused by incorrect
readings, charge the meter immediately when it shows the
" " symbol.
1.3.4Clean the meter with a damp cloth and mild detergent. Do not
use abrasives or solvents.
1.3.5Turn off the power when the meter is not in use.
1.3.6The meter cannot be charged for a long time. After being fully
charged, unplug the Type-c line charging cable.
2describe
2.1 Part name
Non-contact voltage detection
sensing area
Type-c charging cable port
LCD touch button
Display
Input socket
The power button
VΩ
2
describe
2.2 LCD touch display
AC & DC
Connection/Disconnection indication
Automatic range mode
Automatic shutdown indication
Low battery
Reading hold state
Volt (voltage), ampere (current)
Ohm kilohm and megohm (resistance)
Hertz
DIODE
Microfarad
Centigrade and Fahrenheit
The line of fire test
3
Technical indicators
3. Technical indicators
Gauges shall be specified for a period of one year, between
18°C and 28°C, with relative humidity less than Recalibration in
75% condition.
3.1 Comprehensive Indicators
1.Automatic selection of measuring function and range.
2.Full range overload protection.
3.The maximum voltage between the measuring end and the
ground is 1000V DC or 750V AC
4.Working height: maximum 2000m
5.Display: LCD
6.The maximum display value is 9999 digits.
7.Polarity indication: automatic indication, '-' indicates negative
polarity.
8.Superview display: 'OL' or '-OL'.
9.Sampling time: about 3 times/second
10.Unit display: with function, electric quantity unit display.
11.Automatic shutdown time: 15 minutes
12.Power supply: 3.7V/600mAh lithium battery
13.Battery undervoltage indicator: LCD symbol.
14.Temperature coefficient: less than 0.1X accuracy /°C
15.Operating temperature: 0° C-40 ° C
16.Storage temperature: -10°C- 50°C
4
Technical indicators
3.2Technical indicators
3.2.1Alternating current
Range Resolution Accuracy
999.9mA
± (1.0% reading + 8 digits)
10A
-Minimum input current 20mA AC current
-Maximum input current 10A AC current
-Frequency range: 40HZ-1000Hz;
3.2.2Dc current
Range Resolution Accuracy
999.9mA
± (1.0% reading +5 digits)
10A
-Minimum input current 20mA DC current
-Maximum input current 10A DC current
3.2.3Dc voltage
Range Resolution Accuracy
999.9mV
9.999V ± ( 0.5% reading + 3 digits)
99.99V
999.9V
-In AUTO mode, the input voltage ranges from 0.5V to 1000V DC
-The input voltage ranges from 0.0mV to 999.9mV DC in mV mode
Note:
In the mV range, with the input open, the meter may have a
pulsating reading, This is normal because of the high sensitivity
of the meter when connected to the pen When you measure
the circuit, you get a true measurement.
5
Technical indicators
3.2.4Ac voltage
Range Resolution Accuracy
999.9mV
9.999V
±( 0.8% reading + 5 digits)
99.99V
999.9V
-In AUTO mode, the input voltage ranges from 0.5V to 1000V AC
-In mV mode, the input voltage ranges from 0.0mV to 999.9mV AC
-Frequency rate pattern: 40HZ-1000Hz
3.2.5Line connection test
Range Resolution explain
If resistance of the line be ing measured is
less than 50 Ω , buzzer in instrument
will make continuous alarming sounds.
-Overload protection: 600V DC or AC (valid value)
3.2.6 Resistance
Range Resolution Accuracy
999.9Ω
9.999KΩ
99.99KΩ
± ( 0.8% reading + 3 digits)
999.9KΩ
9.999MΩ
99.99MΩ
-Input protection: maximum 600V DC or AC valid value.
6
Technical indicators
3.2.7Diode
Range Resolution Accuracy
-Input protection: the maximum value is 600V DC or AC.
3.2.8Capacitance
Range Resolution Accuracy
9.999nF reading digits
999.9nF
9.999uF
99.99uF
±( 2.5% reading+20 digits)
999.9uF
9.999mF
99.99mF
-Input protection: the maximum value is 600V DC or AC.
3.2.9The temperature
Range Resolution Accuracy
± (1% reading +3℃)
± (1% reading +3℉)
-Accuracy does not include error of thermocouple probe.
-Overload protection 250V DC or AC (valid value)
7
Operational guidelines
3.2.10Frequency
3.2 .10.1 Pass grade A:
Range Resolution Accuracy
reading digits
-Measuring range: 40HZ-1000Hz
-Input signal pattern: > 3A AC current (Effective Value)
3.2.10.2 Through V gear:
Range Resolution Accuracy
reading digits
-Measuring range: 40HZ-1000Hz
-Input signal pattern: > 0.8V AC current (Effective value)
4.Operational guidelines
4.1Reading hold
In the process of measurement, if you need to hold the reading,
touch the key, the display value will be locked, touch the key
again, can remove the reading hold.
4.2Flashlight
Touch the button to turn on the lighting function of the
flashlight barrel, which will be automatically turned off after about
1 minute.。
4.3Automatic shutdown
4.3.1If there is no operation within 15 minutes after startup, the
meter will enter hibernation state and automatically shut down
to save electric energy. The buzzer has one sound one minute
before the shutdown.
4.3.2After automatic shutdown, press the power button to wake up
the instrument to work.
8
Operational guidelines
4.4Preparation before measurement
4.4.1 Tap the power source button to power on. If the battery
voltage is low (about 2.8V), the monitor will display " "
symbol, then should charge. Press the power button again
to turn it off.
4.4.2 When the instrument is not measured, the instrument enters
the automatic scanning state and the instrument displays
4. 5 Non-contact Voltage Detection (NCV)
4.5.1Touch the NCV key to enable the NCV function, and the
instrument displays NCV
4.5.2The NCV sensor is placed close to the measured wire, and the
instrument can detect whether the measured wire is gt. 90V
ac voltage. When the meter detects the ac power, the meter
buzzer alarms and the LCD analog bar displays the induction
intensity.
VΩ
Note:
1) Voltage may still exist even if there is no alarm indication. Do not
rely on contactless voltage detectors To see if there's a voltage
on the wire. Probe operation may depend on socket design,
insulation thickness Different types of factors such as the impact.
2) In NCV mode, the meter does not measure voltage, resistance,
and current at the same tim
9
Operational guidelines
4.6Ac-dc current measurement
4.6.1Insert the stylus into the 10A current input hole, and the
current input terminal is inserted into the stylus automatic
identification function on on Automatically enters the default
DC current gear of the current gear system. Remove the 10A
current input hole and the stylus automatically jumps out of the
current gear and enters the AUTO measurement mode
Note:
When the meter is in diode
gear, the pen is inserted into the
10A input hole and the meter will
not enter current gear.
4.6.2Connect the pen into the
measured signal, when the
measured signal > The meter
will display the current measured
value at 20mA. When measured
ac signal > At 3000rnA, the VΩ
instrument pair displays the AC
current frequency.
4.7Dc voltage measurement
Insert the black stylus into the COM jack and the red stylus into
the V jack. Connect the markers together The voltage source or
load ends are measured and read on the LCD.
AUTO automatic measurement mode:
when the measured signal > At 0.5V, the meter will display
the current measured DC voltage value. When the signal being
measured. At 0.5V, the meter will default to the resistance value
and display the internal resistance value of the measured signal.
Touch the " " key to switch to the DC mV voltage level.
Connect the pen to both ends of the voltage source or load for
measurement, reading on the LCD.
10
Operational guidelines
DC mV measurement mode:
when the measured signal < At
600.0mV, the meter will display
the current measured DC voltage
value. When the measured signal
> The meter will display "OL"
at 999.9mV.
VΩ
4.8Ac voltage measurement
Insert the black stylus into the COM jack and the red stylus into
the V jack. Connect the pen to both ends of the voltage source or
load for measurement, reading on the LCD.
AUTO automatic measurement mode:
when the measured signal > At 0.5V, the meter will display the
current measured AC voltage value. When the signal being
measured. 0.5 VInstrument will
default to the resistance value,
display the measured signal internal
resistance value.
Touch the " " key to switch to ac
mV voltage. Connect the pen to both
ends of the voltage source or load
for measurement, reading on the LCD.
AC mV measurement mode:
when the measured signal < At
600.0mV, the meter will display the VΩ
current measured AC voltage value.
When the signal is measured>
The meter will display "OL" at 999.9mV."
11
Operational guidelines
4.9 Resistance measurement
warning
When measuring the impedance on the line, make sure
that the circuit is disconnected and the capacitor on the
circuit discharges completely.
4.9.1 Insert the black stylus into the COM jack and the red stylus
into the " Ω " jack.
4.9.2 AUTO AUTO range tap. At this time, the meter is in automatic
scanning state.
4.9.3 Connect the pen to the measured resistance or both ends of
the line for measurement, reading in the LCD display.
Note:
1) When the input is open, the LCD will display the "OL"
overrange status.
2) If the resistance being measured is higher than 10MΩ,
the meter may take several seconds to stabilize the
reading, which is normal for high resistance readings.
VΩ
12
Operational guidelines
4 .10 The line of fire test
4.10.1Touch the "Live" key to switch to LIVE gear, the instrument
shows Live
4.10.2Connect the red test pen to the Live input socket. Insert the
single pen into the power socket L jack or close to the live
wire. If the instrument detects ac voltage, it will display
voltage signal strength according to the detected signal
strength, and the buzzer will give alarm sound of different
intensity.
4 .11Capacitance measurement
4.11.1Touch the " " key to switch capacitor gear.
4.11.2Use the other end of the test pen to measure the capacitance
value of the capacitor to be measured and read the measured
value from the LCD.
Note: 1)When measuring large capacitors, a fixed time is
required to stabilize readings.
2)When measuring polar capacitors, pay attention to
the corresponding polarity to avoid damaging the
instrument.
4 .12Temperature measurement
Touch "SEL" key to switch TEMP file, thermocouple red plug
into the end of °C /°F, black plug into the COM jack. When the
reading is stable, the temperature can be read directly from the
display screen.
Note:
The maximum measurement temperature of type K
thermocouple is 250°, instantaneous measurement up to 300°C.
4.13Diode test
4.13.1 Insert the black stylus into the COM jack and the red stylus
into the " " jack.
4.13.2 Touch SEL key to switch to diode position.
4.13.3 Connect the red stylus to the diode anode and the black
stylus to the diode cathode for testing.
13
Operational guidelines
4.13.4 Read on the LCD.
Note:
The meter displays an
approximation of the forward
voltage drop of the diode.
If the pen is connected
inversely or the pen is open,
the LCD displays "OL".
4 .14Line connection test
VΩ
warning
Danger of electric shock.
When testing the circuit on and
off, make sure the circuit is
disconnected and the capacitor
on the circuit is completely
discharged
4.14.1Insert the black marker into
the COM jack and the red
marker into the " " jack.
4.14.2Touch the "SEL" key to switch
to the line connection test state.
4.14.3Connect the pen to both ends VΩ
of the line for measurement.
If the resistance of the measured
line is less than 50 Ω, the buzzer inside the meter will sound.
Note:
If the meter is open or the resistance of the measured
line is greater than 999.9Ω, "OL" is displayed.
14
Maintenance
5.Maintenance
5.1 Type - c line charging
warning
During the charging process, the meter pen should be
removed from the measuring circuit to avoid the danger of
lectric shock.
If the " " symbol appears, it means to charge.
5.2Replace the test pen
warning
When replacing the stylus, you must replace the same or
the same grade of the table. Pen meter must be in good
condition, pen grade: 1000V 10A.
Note:
If the insulation layer of the pen is damaged, such as
exposed wire of the wire, the pen must be replaced.
6.The attachment
1)Test the pen Grade: 1000V 10A A pair of
2)Directions for use A copy of the
3)thermocouple a
4)Charging line Type-c a