Scanner Specifications Reference

Reference parameters for clinical CT scanners used in BasisSimulator.jl verification. These were previously implemented as factory functions and have been moved here for reference.

GE Revolution Apex Elite (Energy-Integrating)

FDA 510(k): K213715

Parameter Value
SID 626.0 mm
SDD 1097.0 mm
Detector Rows × Cols 256 × 832
Row Size (iso) 0.625 mm
Col Size (iso) 1.053 mm
Z Coverage 160 mm
Detector Material Lumex (garnet scintillator)
Detector Depth 3.0 mm
Fill Factor 0.90
Focal Spot (small) 1.0 × 0.7 mm
Focal Spot (large) 1.6 × 1.2 mm
Target Angle 10°
Flat Filter 2.5 mm Al
Bowtie large_body
Supported kVp 70, 80, 100, 120, 140
Max mA 1300
Min Rotation 0.23 s
Bore 800 mm
Scan Diameter 500 mm
Detection Gain 15.0 e⁻/keV
Electronic Noise 5000 e⁻

Siemens NAEOTOM Alpha (Photon-Counting)

FDA 510(k): K201501

Parameter Standard (2×2) UHR (1×1)
SID 595.0 mm 595.0 mm
SDD 1085.5 mm 1085.5 mm
Subpixel (iso) 0.300 × 0.176 mm 0.150 × 0.176 mm
Binning Factor 2 1
Pixel Size at iso (marketed) 0.4 mm row × 0.5 mm col 0.2 mm row × 0.25 mm col
Detector Rows 144 120
Z-coverage at iso (marketed) 57.6 mm (144 × 0.4 mm) 24.0 mm (120 × 0.2 mm)
Detector Material CdTe CdTe
Detector Depth 1.6 mm 1.6 mm
Fill Factor 0.95 0.95
Energy Thresholds 20, 35, 55, 70 keV 20, 35, 55, 70 keV
Energy Resolution 10 keV FWHM 10 keV FWHM
Charge Sharing FWHM 0.08 mm 0.08 mm
Dead Time 5.0 ns 5.0 ns
Focal Spot 1.0 × 1.0 mm 0.4 × 0.5 mm
Target Angle
Flat Filter 2.5 mm Al 2.5 mm Al
Supported kVp 70, 90, 120, 140 70, 90, 120, 140
Min Rotation 0.25 s 0.25 s
Bore 820 mm 820 mm
Scan Diameter 500 mm 500 mm

Dual-source geometry: Two source/detector pairs at ~90° angular offset around the gantry — both cover the same Z extent. The second source is for temporal resolution (66 ms quarter-rotation imaging) and high-pitch (3.2) helical Flash mode, not for extended Z coverage. Max single-rotation axial Z coverage is 57.6 mm in standard mode; Flash mode achieves ~120 mm of scan range in ~160 ms via table translation, not a wider detector.

For simulator scenarios that want to approximate Flash-mode coverage as a single axial rotation (training-data generation, plaque-volume scans), use the extended_collimation kwarg on CTGeometry(scanner; collimation_mm=..., extended_collimation=true) — see that function's docstring.

Note on row pitch: The 0.4 mm marketed pitch is what Siemens publishes (Petersilka et al. 2021; vendor planning docs) and is the canonical number used throughout the literature. A strict geometric derivation from the 0.176 mm subpixel and 1.825× magnification would give 0.176 × 2 = 0.352 mm — a ~13% underestimate. Internally BasisSim adopts the published 0.4 mm so detector_rows × detector_row_size matches the 57.6 mm spec.

Canon Aquilion ONE (Legacy Reference)

Parameter Value
SAD 600 mm
SDD 1000 mm
Pixel Pitch (iso) 0.5 mm
Detector Type Energy-integrating